INTRODUCTION:
The Circulating Bath with chiller is extensively utilized across a range of fields including Petroleum, Chemical, Electronic, Instrumentation, Physics, Chemistry, Bioengineering, Medical and Health, Life Sciences, and Food Industries. It is essential for property testing and chemical analysis, serving research departments, universities, and quality inspection and production departments in enterprises.
WARNING:
- Electrical Hazard: To prevent electric shock, do not operate this equipment in wet conditions. Ensure that all electrical installations comply with national and local codes. Disconnect power before servicing.
- Chemical Exposure: Use appropriate protective equipment when handling chemicals that are being cooled or heated in the chiller to avoid skin and eye contact.
- Pressure Build-up: Do not obstruct the flow of coolant in the circulation system. Blockages can cause pressure to build up, leading to equipment damage or personal injury.
- Water Quality: Use only clean, deionized water or other specified coolants to avoid scale buildup and corrosion, which could lead to leaks and system failure
- Improper Use: This chiller is designed for specific cooling tasks. Using the Circulating Bath for applications outside its intended use may compromise safety and equipment performance
- Ventilation: Ensure adequate ventilation around the Circulating Bath to prevent overheating and potential failure of the cooling system.
PRECAUTIONS FOR USE:
- Before operation, ensure that the tank is filled with a liquid medium; operating without liquid is strictly prohibited. Note: the liquid should not contain acids or alkalis.
- The power supply should have a capacity greater than the total power of the instrument, and it must be properly grounded. Use a voltage according to the specifications listed in the technical parameter table.
- Position the instrument in a well-ventilated and dry area, ensuring at least 300 mm of clearance from any obstacles at the back and sides to facilitate good heat dissipation.
- After use, turn all switches to the off position, unplug the power cord, open the drain valve to remove the medium from the tank, and keep the tank dry and clean
FEATURES:
- Refrigeration System: The unit uses a low-noise, air-cooled, fully enclosed compressor, which offers rapid cooling and excellent temperature stability.
- Control System: It incorporates the latest proprietary software and temperature control technologies, featuring a high-standard PT100 and top-quality imported electronic components.
- Circulation System: This system includes both internal and external circulation. External circulation allows the constant temperature liquid from the tank to be directed outside, creating a secondary temperature-controlled environment or serving as a cold source to cool experimental containers outside the machine. Internal circulation ensures uniform and stable liquid temperatures within the tank.
- Protection System: The refrigeration system is equipped with overheating and overcurrent protections, while the control system includes over-temperature alarms, settings for temperature limits, and automatic disconnection of the load in the event of excessive temperatures.
- Display and Interface: Features a high-end, large-screen LCD control instrument with a high resolution of 0.01°C, offering simple operational functions and robust temperature stability. PID control allows for automatic tuning or manual adjustment.
- Temperature Accuracy: Includes precise temperature correction up to 0.01°C, with maximum temperature fluctuations is ±0.02°C.
SPECIFICATIONS:
MODEL | IG-62CB |
---|---|
Temperature Range | -20ºC to 100ºC |
Temperature Display resolution | ± 0.1ºC |
Temperature Fluctuation | ± 0.2ºC |
Temperature Control | Microprocessor PID |
Display | LCD |
Refrigeration | Air Cooled refrigeration through compressor |
Refrigerant | R-134a |
Pump circulation | Internal & External |
Pump flow rate | 6 L /min |
Bath Tank capacity | 6L |
Tank size (Lx W x D) | 180 x 290 x150 mm |
Bath Opening | 180 x 140 mm |
Bath depth | 150 mm |
Power Requirements | 220VAC/50Hz |
Packing Dimensions (LxWxH) | 530 x 450 x 800 mm |
Weight(LxWxH) | 44 kg |
Heating Capacity | 2000 W |